Skip to content

Public FDA record

Z-2172-2021: GlideScope Go Monitors

Recall record Z-2172-2021 for Verathon, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2172-2021
Company in FDA record
Verathon, Inc.
FDA record creation date
Jul 30, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Handheld video monitor failed to meet the labeled IP67 rating which may allow fluid ingress into the monitor resulting in loss of functionality.

Source facts

Action
On 07/01/2021, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification Letter via FedEx to customers informing them that the Recalling Firm has become aware of a variation in material thickness of a micro USB mounting plate can lead to insufficient sealing, resulting in the device failing to meet the labeled IP67 rating allowing fluid ingress and impact functionality of the monitor. Customers are instructed to: 1. Perform a visual inspection of monitor for fluid ingress, referring to figures 1 and 2 of the Notification Letter for visual examples of the failure to meet the IP67 rating. 2. Based on the visual inspection of their device(s), do one of the following: -If their Monitor has no fluid ingress or they do not wish to return their affected product, complete the attached Recall Response Form by indicating no or 0 in the text box and return the form by email to CSNotifications@verathon.com. Or -If their Monitor has fluid ingress and they would like these units serviced (similar to Figures 1 and 2) complete the attached Recall Response Form. They will need to indicate, by number, the count of Monitors that require service in the text box and return the form by email to CSNotifications@verathon.com. The Recalling Firm's Customer Care will contact them to make arrangements to replace their monitor. The firm is implementing a change in the design for the monitor micro-USB mounting plate to reduce the variation in the thickness in order to meet the IP67 rating. Any questions - contact the Recalling Firm's Representative or Customer Care at CSNotifications@verathon.com
Initiated
2021-07-01
Root cause
Device Design
Distribution
U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, Puerto Rico, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY. O.U.S.: Australia, Brunei Darussalam, Canada, Czech Republic, Denmark, Estonia, Falkland Island (Malvinas), France, Germany, Greece, Hong Kong, Iceland, Ireland, Israel, Italy, Kuwait, Luxembourg, Malaysia, New Zealand, Oman, Papua New Guinea, Philippines, Portugal, Qatar, Reunion, Romania, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, The Netherlands, United Arab Emerita, and United Kingdom.
Recall number
Z-2172-2021
Recall status
Open, Classified
CDRH record ID
188141
FDA record created
2021-07-30
Recall event number
88214

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗