Z-2173-2013: KIT cobas 4800 HPV AMP/DET 240T / 960T US-IVD, CE-IVD Roche Molecular System, Inc. 1080 US highway 202 So

Recall record Z-2173-2013 for Roche Molecular Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2173-2013
Company in FDA record
Roche Molecular Systems, Inc.
FDA classification date
Sep 10, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During an internal investigation, a formulation error that occurred was identified during the manufacture of the cobas 4800 HPV Master Mix reagent that was packaged in four (4) cobas 4800 AMP/DET kits, US- and CE-IVD (240 and 960 test kit configurations). Examination of the batch production record of the material revealed there was an error in the tare weight used for the carboy during the fina

Source facts

Action
Roche Molecular Diagnostics sent an Urgent Medical Device Correction letter dated July 25, 2013, to all affected customers via UPS return receipt. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to: 1. Immediately discontinue use of the product lots listed above. 2. Discard affected product according to local guidelines. 3. To receive replacement inventory, complete the entire attached faxback form and fax it to 1-866-918-8741. 4. The faxback form must be completed and returned even if you do not have affected inventory. 5. File this Urgent Medical Device Correction (UMDC) for future reference. Please contact Roche Molecular Diagnostics Technical Support, 24 hours a day, seven days a week at 1-800-526-1247, if you have questions about the information contained in this UMDC.
Initiated
2013-07-25
Root cause
Device Design
Terminated
2015-08-06
Distribution
Worldwide Distribution - USA (nationwide) including the countries of Austria, Australia, Belgium, Brazil, Czech Rep, Canada, Croatia, Columbia, Denmark, Dominican Republic, Germany, Ecuador, France, Japan, Greenland, Mexico, Italy, New Zealand, Netherlands, Peru, Norway, Singapore, Romania, South Africa, Spain, South Korea, Sweden, Switzerland and UK.
Recall number
Z-2173-2013
Recall status
Terminated
CDRH record ID
121300
FDA classified
2013-09-10
Recall event number
66036

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗