Open original FDA record ↗Recall record
- FDA identifier
- Z-2175-2008
- Company in FDA record
- Abbott Laboratories
- FDA classification date
- Sep 16, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
An increase in complaints was noted for AxSYM Rubella IgG assay calibration failures due to error codes related to elevated Calibrator A (or Master Calibrator 1) rates too high.
Source facts
- Action
- Consignees were sent a Product Correction-Immediate Action Required letter on 6/2/08. The letter informed users of calibration failures due to elevates Calibrator A or Master Calibrator 1 rates when attempting to calibrate the AxSYM Rubella IgG assay and that they must clarify the Rubella IgG Calibrator A or the Rubella IgG Master Calibrator 1 and the Rubella IgG Negative Control by centrifugation at >/= 10,000 x g for 10 minutes prior to testing. Users were requested to return a customer reply letter acknowledging receipt of the letter. For additional information, contact Abbott Labs at 1-877-4ABBOTT.
- Initiated
- 2008-06-02
- Root cause
- Device Design
- Terminated
- 2008-11-05
- Distribution
- Worldwide Distribution including USA, territory of Puerto Rico, and countries of Australia, Bahamas, Canada, Chile, Colombia, Costa Rica, Curacao, Dominican Republic, Ecuador, El Salvador, Germany, Guatemala, Honduras, Hong Kong, Jamaica, Mexico, Peru, New Zealand, Panama, Singapore, South Korea, Thailand, Turks & Caicos and Uruguay.
- Recall number
- Z-2175-2008
- Recall status
- Terminated
- CDRH record ID
- 71762
- FDA classified
- 2008-09-16
- Recall event number
- 48626