Open original FDA record ↗Recall record
- FDA identifier
- Z-2175-2013
- Company in FDA record
- B Braun Medical, Inc.
- FDA classification date
- Sep 10, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
B. Braun Medical Inc. is informing our customers of a field correction previously performed to address the loss of the drug library in the Outlook Pump Models 620-100 and 620-200. This could have occurred after periods of lost DC (battery) power or out of tolerance low power. This could have led to a delay in therapy, or to delivery of an over or under-dose of a therapy if the user was to rely on
Source facts
- Action
- The firm, B. Braun Medical Inc., sent an "URGENT: MEDICAL DEVICE FIELD CORRECTION" letter dated December 05, 2012 to its customers. The letter describes the product, problem and actions taken. The firm finished performing "field upgrades" on March 20, 2010 for this issue. Should you require any additional information, please do not hesitate to contact B. Braun Medical Inc., Infusion Systems Customer Service at 1-800-627-7867 (627-PUMP).
- Initiated
- 2012-12-05
- Root cause
- Component design/selection
- Terminated
- 2014-10-24
- Distribution
- Worldwide Disribution: USA (nationwide) and country of: Canada.
- Recall number
- Z-2175-2013
- Recall status
- Terminated
- CDRH record ID
- 115430
- FDA classified
- 2013-09-10
- Recall event number
- 64150