Open original FDA record ↗Recall record
- FDA identifier
- Z-2175-2014
- Company in FDA record
- AGFA Healthcare Corp.
- FDA classification date
- Aug 10, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Some customers reported that while using the DX-D600 in room or exam preparation, the Overhead Tube Crane (OTC) showed increased velocity and uncontrolled longitudinal movement.
Source facts
- Action
- AGFA HealthCare sent an URGENT FIELD SAFETY NOTICE letter dated July 2, 2014 to affected customers. The letter described the affected product, problem and actions to be taken. Customers were instructed to complete and return the Acknowledgment form, via FAX-Back or email. A Mandatory Service Bulletin DIS012.14E/No. 32, was released June 30, 2014, to implement the mandatory upgrade of full automatic DX-D600 systems to version 3.6 to prevent unexpected system movements. This Mandatory Service Bulletin was issued to Agfa DR Specialists to perform the service. A 2nd release will be implemented to upgrade semi-automatic systems. For questions contact your local AGFA HealthCare organization 1-877-777-2432.
- Initiated
- 2014-07-02
- Root cause
- Device Design
- Terminated
- 2015-06-04
- Distribution
- Worldwide Distribution - US Natinwide in the state of GA, IN, KS, KY, MA, NJ, SC, VA, DC and the country of Canada
- Recall number
- Z-2175-2014
- Recall status
- Terminated
- CDRH record ID
- 128600
- FDA classified
- 2014-08-10
- Recall event number
- 68752