The patient support head holder could contact the back ISO shelter wall during manual or motorized motion through the gantry bore. This may result in a serious injury that would require medical intervention to preclude permanent impairment in certain scenarios.
Source facts
Action
Urgent Field Safety Notices dated 3/28/18 were distributed to customers. The letters instructed customers to perform the following actions:
Please review the following information with all members of your staff who need to be aware of the contents of this communication. It is important to understand the implications of this communication.
Please retain a copy with the equipment Instruction for Use.
As outlined in the Instructions for Use, always monitor the patient during all
movements (manual or motorized) of the patient support. If you recognize an impending collision with the back shelter wall, halt the motion or activate the system E-stop.
Mandatory Field Change Order (FCO) 72800686 which will add additional
warning labels specific to this issue to the system. The FCO will be implemented free of charge.
Initiated
2018-03-30
Root cause
Device Design
Terminated
2020-04-14
Distribution
Worldwide distribution including US states of CA, MD, NV, OH, TX, and WI, Afghanistan, Canada, Germany, Iraq, Italy, Korea, Kuwait, and Lebanon.
Recall number
Z-2176-2018
Recall status
Terminated
CDRH record ID
164549
FDA record created
2018-06-12
Recall event number
80094