Skip to content

Public FDA record

Z-2176-2024: The VASOVIEW HemoPro Endoscopic Vessel Harvesting Systems are designed for use in conjunction with the 7

Recall record Z-2176-2024 for Maquet Cardiovascular, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2176-2024
Company in FDA record
Maquet Cardiovascular, LLC
FDA classification date
Jun 24, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for there to be a fluid ingress into the handle of the harvesting tool, which may lead to the device not providing cautery or providing cautery when not intended.

Source facts

Action
Consignees were mailed an URGENT MEDICAL DEVICE - CORRECTION notice, dated 5/17/24. In the notice consignees are instructed to make users aware of the additional safety information provided in the notification by forwarding the notice to users in their facility and locations where devices are stored. Additionally, consignees are asked to return the provided response form by email to VVHemoPro2024.act@getinge.com or by fax to 800-892-0487. Getinge will be updating the Instructions for Use (IFU) of devices to include the additional safety information related to fluid ingress. Questions can be directed to Customer Support at 888-880-2874, Monday through Friday from 6:00 AM to 5:00 PM PST. UPDATE: On 07/11/2025 an updated Urgent Medical Device Correction Update letter was posted on Getinge.com and sent to consignees via letter. The updated communication informs users that the Instructions for Use (IFU) for the Vasoview Hemopro 2 EVH System has been revised to include warnings and precautions regarding fluid ingress into the handle of the Harvesting Tool. Consignees were instructed to review the updated IFU, view the training video linked in the letter, complete and return the Medical Device Correction Response form, and notify customers if product was further distributed.
Initiated
2024-05-17
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide and the countries of Australia Brazil, Canada, China, EEA, India, Korea, New Zealand, Southern Africa, Taiwan, Thailand, Turkey, & United Kingdom.
Recall number
Z-2176-2024
Recall status
Open, Classified
CDRH record ID
208047
FDA classified
2024-06-24
Recall event number
94700

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗