Open original FDA record ↗Recall record
- FDA identifier
- Z-2177-2025
- Company in FDA record
- Steris Corporation
- FDA classification date
- Jul 25, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The DFPM yoke assembly ("yoke assembly") may detach from the spring arm. Should the yoke assembly detach from the spring arm, the yoke assembly would not be expected to fully detach but rather hang from the wiring within the yoke assembly.
Source facts
- Action
- Steris issued an URGENT MEDICAL DEVICE RECALL NOTICE dated 06/17/2025 to its consignees on 06/23/2025 disseminated via FedEx. The notice explained the issue, potential risk, and requested the following: User Action: Please ensure the following steps are completed: 1. Review the revised Preventive Maintenance Manual and Interval Based Check List language depending on which HarmonyAIR Surgical Lighting System model(s) are in use at your facility (A-Series and/or E-Series) (please reference Attachments B-E of the letter for maintenance instructions). 2. Immediately complete the maintenance activities as instructed, and then again annually going forward. 3. Complete the Medical Device Recall Response Form included with this letter. 4. Return the completed Response Form via email to: Regulatory_Compliance@STERIS.com, or via fax to 440-392-8963. 5. Distributors: If you have further distributed this product, please identify your Customer(s), and notify them of this recall.
- Initiated
- 2025-06-23
- Root cause
- Process control
- Distribution
- Worldwide - US Nationwide and Puerto Rico and the country of Canada.
- Recall number
- Z-2177-2025
- Recall status
- Open, Classified
- CDRH record ID
- 214408
- FDA classified
- 2025-07-25
- Recall event number
- 97068