Skip to content

Public FDA record

Z-2179-2017: Table Patient Step on RFX/SFX, Legacy, and Precision 500D systems

Recall record Z-2179-2017 for GE Healthcare, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2179-2017
Company in FDA record
GE Healthcare, LLC
FDA classification date
May 22, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Reported incidents of a patient step detaching from the table. A fall from a patient step detaching while in use could result in an injury to a patient or operator.

Source facts

Action
Consignees were sent on 2/6/2017 a GE Healthcare "Urgent Medical Device Correction" letter GEHC Ref# FMI 10908 dated February 3rd, 2017. The letter was addressed to Director of Radiology, Risk Manager/Hospital Administrator & Director of Biomedical Engineering. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact consignee to arrange for the correction.. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Initiated
2017-02-06
Root cause
Device Design
Terminated
2018-03-02
Distribution
US: Nationwide including DC and Puerto Rico. OUS: Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Egypt, France, Germany, Greece, Honduras, Hungary, Indonesia, Ireland, Israel, Italy, Jamaica, Jordan, Kuwait, Libya, Malaysia, Malta, Mexico, Panama, Portugal, Russia, Saudi Arabia, Saudi Arabia, Slovakia, Spain, Thailand, United Arab Emirates, United Kingdom, Vietnam.
Recall number
Z-2179-2017
Recall status
Terminated
CDRH record ID
153410
FDA classified
2017-05-22
Recall event number
76517

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗