Open original FDA record ↗Recall record
- FDA identifier
- Z-2179-2018
- Company in FDA record
- Datascope Corporation
- FDA record creation date
- Oct 11, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a potential for the over-pressure safety (OPS) valve in the tubing kits to not allow flow through.
Source facts
- Action
- On 7/27/17, an Urgent Safety Alert was issued to customers of the Cardiovascular Procedure Kit with LN130B Over-pressure Safety Valves. The safety alert instructed customers to perform the following: Please examine your inventory immediately to determine if you have unused kits of the batch numbers listed. If so, please follow the recommended actions outlined in the attached Terumo Cardiovascular Systems Urgent Safety Alert. It is possible that custom tubing kits with the affected valves may have already been used without issue. Customers were instructed to test the valve prior to use. The notice was sent to customers again on 10/3/17 and 12/12/17. On 1/11/18, Urgent Medical Device Recall letters were sent to customers. The letters included the following instructions: Terumo CVS now recommends removal of any remaining product from your inventory. A review of our records indicates that you have received the following Maquet/Getinge Custom Tubing Kit batch numbers affected by the Terumo Cardiovascular Systems Medical Device Recall (refer to information provided in the table below). Please examine your inventory immediately to determine if you have unused custom tubing kits of the batch numbers listed. If so, please remove the affected products, quarantine them and follow the instructions for returning the product to Maquet/Getinge for full credit. It is possible that custom tubing kits with the affected valves may have already been used without issue. The Terumo Cardiovascular Systems Medical Device Recall affects the Custom Tubing Kit with LN130B Over-pressure Safety Valve distributed by Maquet/Getinge having lot/batch numbers listed above; no other product lot/batch numbers that were delivered to you are affected. Instructions for returning affected product: 1. Review this notice and Terumos Medical Device Recall Notice (attached). 2. Ensure that all users (i.e. perfusionist, operating room staff and biomedical staff) receive notice
- Initiated
- 2017-07-27
- Root cause
- Nonconforming Material/Component
- Terminated
- 2020-02-07
- Distribution
- The products were distributed to the following US states: IA and FL.
- Recall number
- Z-2179-2018
- Recall status
- Terminated
- CDRH record ID
- 164457
- FDA record created
- 2018-10-11
- Recall event number
- 80081