Z-2180-2008: Impax 6.x Clients including: 6.3 all versions below 6.3.1 SU5, 6.2 all versions below 6.2.1 SU14 and 6.0

Recall record Z-2180-2008 for AGFA Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2180-2008
Company in FDA record
AGFA Corp.
FDA classification date
Sep 4, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The IMPAX client failed to "refresh" the image area upon receiving a study retrieved from archive. The system also erroneously displays this study retrieved from archive as the "active" study.

Source facts

Action
An "Urgent Safety Notice" letter was sent to all the sites using potentially affected units on/about June 4, 2008. Dealers were notified via letter on June 9, 2008. The letter requests that all users be reminded that Agfa recommends the system is configured so that demographic information is not removed from the image viewing area. They will be provided with the appropriate Service Update, based on the current version of IMPAX Client 6.x in use at their site. They were advised to distribute the information within their facility to all those who need to be aware of it. A FAX Back form was requested with each letter to ensure someone at the account acknowledges the information was received and understood. If a FAX Back form is not received, the account will be contacted by telephone. Contact AGFA at 1-877-777-2432, prompts 1, 2, 1.
Initiated
2008-06-04
Root cause
Software design
Terminated
2009-10-09
Distribution
U.S., Canada, Europe, Asia Pacific and Latin America.
Recall number
Z-2180-2008
Recall status
Terminated
CDRH record ID
71764
FDA classified
2008-09-04
Recall event number
48632

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗