- FDA identifier
- Z-2180-2011
- Company in FDA record
- Digisonics, Inc
- FDA classification date
- May 10, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Data from one patient's study had the potential for being put into another patient's report.
Source facts
- Action
- Digisonics notified end users via e-mail January 20, 2009 and January 21, 2009, that Data from one patient's study was being put into another patient's report. The problem occurred when a System bug was found in DigiView Release 3.6.4.4, and higher versions in which under certain circumstances (specific workflow) a failure to switch to a patient's images occurs. Two software patch versions will be made available to sites on or after February 5, 2009, to correct the reported problem. If you have any further questions, please contact Digisonics Customer Support at (800 ) 940-3240.
- Initiated
- 2009-01-20
- Root cause
- Device Design
- Terminated
- 2011-05-31
- Distribution
- Worldwide Distribution -- USA including States of : OH, NC, LA, TX, CA, MO, NY, MD, MA, DC, IA, AZ, PA, MI, NC, FL, IN, SC, ID, MN, AR, OK, IL, CT, and countries of Canada, Lebanon, and New Zealand.
- Recall number
- Z-2180-2011
- Recall status
- Terminated
- CDRH record ID
- 99262
- FDA classified
- 2011-05-10
- Recall event number
- 58395