The manufacturer of the DOLPHIN inflation device, Perouse Medical, has initiated a recall of the device due to a potential defect in the sterile barrier. Use of affected product could pose a risk of microbial contamination, leading to infection.
Source facts
Action
Customers were sent a Vascular Solutions "Urgent Medical Device Removal letter" dated April 12, 2017. The letter described the problem and the product involved in the recall. Advised consignees to remove and secure all affected product and to complete and return the "VSI Account Inventory Form". For questions contact your local Vascular Solutions Sales Representative. For further questions, please call (763) 656-4300.
Initiated
2017-04-12
Root cause
Packaging
Terminated
2018-04-11
Distribution
US Distribution including PR and to the states of : LA, TN, CA, SC, OR, WA, OK, NV, HI, OH, TX.
Recall number
Z-2180-2017
Recall status
Terminated
CDRH record ID
154966
FDA record created
2017-05-31
Recall event number
77095