Open original FDA record ↗Recall record
- FDA identifier
- Z-2181-2012
- Company in FDA record
- Kimberly-Clark Corporation
- FDA classification date
- Aug 9, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
A defect of the "fuse holder" can lead to a Control Unit failure that may cause the device to lose its ability to warm and maintain set temperature and may also result in excessive heat with the fuse assembly that could lead to the generation of smoke.
Source facts
- Action
- Kimberly-Clark sent an "URGENT: FIELD CORRECTIVE ACTION NOTICE" dated April 27, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. A Field Corrective Action Response Form was attached for customers to complete and return. Contact the Field Corrective Action Coordinator at 770-587-7400 for questions regarding this notice.
- Initiated
- 2012-04-27
- Root cause
- Device Design
- Terminated
- 2012-09-05
- Distribution
- Nationwide Distribution-including the states of AZ, CA, CO, DE, FL, GA, IA, ID, IN, IL, KS, KY, LA, MA, MD, ME, MO, MS, NE, NM, OH, OR, PA, RI, SC, TN, TX, VA, WI and WV.
- Recall number
- Z-2181-2012
- Recall status
- Terminated
- CDRH record ID
- 111119
- FDA classified
- 2012-08-09
- Recall event number
- 62596