Open original FDA record ↗Recall record
- FDA identifier
- Z-2183-2011
- Company in FDA record
- Remote Diagnostic Technologies Ltd.
- FDA classification date
- May 11, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
It was discovered that when a 12 Lead ECG is recorded and monitoring is quickly restarted, the ECG can cease functioning.
Source facts
- Action
- Remote Diagnostic Technologies, Ltd notified its consignees of the problem and the work-around via letter with the subject line "ECG Issue - Field Action Required" on 03/25/2011. The letter stated that customers can avoid the issues by viewing the recorded ECG for at least 13 seconds. Two diagrams were attached with the letter depicting the conditions under which the issue can occur. Customers can continue to use the device normally in all other respects. The letter states that the firm is working on a software modification which will permanently fix the issue and prevent it from occurring. Customers are to contact the Regulatory Affairs and Operations Manager if they have any questions.
- Initiated
- 2011-03-25
- Root cause
- Component change control
- Terminated
- 2011-09-16
- Distribution
- Nationwide Distribution -- KY, MD, VA, and FL.
- Recall number
- Z-2183-2011
- Recall status
- Terminated
- CDRH record ID
- 99439
- FDA classified
- 2011-05-11
- Recall event number
- 58479