Z-2183-2014: VITROS Immunodiagnostic Products CA 19-9 Calibrators and Reagent Packs, Lot 1320 VITROS CA 19-9 Reagen

Recall record Z-2183-2014 for ORTHO-CLINICAL DIAGNOSTICS. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2183-2014
Company in FDA record
ORTHO-CLINICAL DIAGNOSTICS
FDA classification date
Aug 11, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Customers may receive positively biased results when using VITROS Immunodiagnostic Products CA 19-9 Total Reagent Packs, Lot 1320 and VITROS CA 19-9 Calibrators, Lot 1320 on the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System. There is an unexpected observed change in CA 19-9 Calibrators, Lot 1320 calibrator values.

Source facts

Action
Ortho Clinical Diagnostics sent an Urgent Product Correction Notification July 2014 to all affected customers. The Notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to Immediately discontinue using and discard Lots 1320 of VITROS CA 19-9 Calibrators and Reagent Packs. Review previously reported VITROS 25-OH Vitamin D Total results using Lot 1320. Review quality control results used to verify calibration events. Results lower than the target mean may indicate potentially biased assay results. Discuss any concerns regarding previously reported results with the Laboratory Medical Director or with the requesting physician. Complete and return the attached Confirmation of Receipt form. OCD will: credit their account for all VITROS CA 19-9 Calibrators and Reagent Packs that were discarded or ship repalcment product uspon request. Customers were instructed to forward the information in this notification if they have distributed this product outside of their facility. Customers with questions were instructed to contact their Customer Technical Service representative or contact the Customer Technical Service representative at 1-800-421-3311.
Initiated
2014-07-16
Root cause
Other
Terminated
2017-03-06
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to France and Japan
Recall number
Z-2183-2014
Recall status
Terminated
CDRH record ID
128956
FDA classified
2014-08-11
Recall event number
68890

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗