Skip to content

Public FDA record

Z-2184-2024: AXIOM Vertix MD - The Vertix MD Trauma has been specially designed for examining emergency and accident p

Recall record Z-2184-2024 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2184-2024
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA classification date
Jun 25, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Short circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire

Source facts

Action
Siemens issued Urgent Medical Device Correction Letter (XP008/24/S) to the Attention: Director of Diagnostic Imaging and/ or Radiology Laboratory on 5/31/24. Letter states reason for recall, health risk, and action to take: What steps can the user take to avoid the possible risks associated with this issue? It is recommended to switch off the system / Touch Display when it is not in use and is unattended. How will the issue be resolved? To ensure continued safe operation of your system and to provide safeguards for the life of the system, the Siemens Healthineers Service Organization will perform the replacement of the PSU of the Touch Display via UI XP006/24/S beginning in the second quarter (June) of 2024. This replacement will be provided to you free of charge. Based on our investigation you can continue to use your system following the recommendation to switch off the system / Touch Display when it is not in use and is unattended. Our service organization will contact you to arrange a date to perform this corrective action once it is available. Please feel free to contact our Customer Care Center at 1-800-888-7436 for an earlier appointment beginning in June 2024. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Customer Safety Advisory Notice. Please respond within 14 days of receiving this notification.
Initiated
2024-05-31
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution.
Recall number
Z-2184-2024
Recall status
Open, Classified
CDRH record ID
208272
FDA classified
2024-06-25
Recall event number
94777

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗