Open original FDA record ↗Recall record
- FDA identifier
- Z-2186-2020
- Company in FDA record
- Osteomed, LLC
- FDA classification date
- May 29, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Incorrect shelf-life for bone void filler kits; the product may not maintain its setting characteristics for the labeled shelf-life duration.
Source facts
- Action
- On 04/02/2020, an URGENT Medical Device Recall letter was sent to all consignees affected by the recall. OsteoMed has requested that affected customers: (1) review this letter in full, review current inventory, and complete and return the Acknowledgement and Receipt Form via email or fax; and (2) return unused inventory via Federal Express (using OsteMed account)
- Initiated
- 2020-04-02
- Root cause
- Vendor change control
- Terminated
- 2021-05-05
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA. MA, MD, ME. MI, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV and WY and the countries of AUSTRALIA, ITALY, KOREA, MALAYSIA, PERU, SINGAPORE and THAILAND.
- Recall number
- Z-2186-2020
- Recall status
- Terminated
- CDRH record ID
- 181180
- FDA classified
- 2020-05-29
- Recall event number
- 85529