Open original FDA record ↗Recall record
- FDA identifier
- Z-2188-2016
- Company in FDA record
- Toshiba American Medical Systems Inc
- FDA classification date
- Jul 18, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
During a procedure the Peak Skin Dose (PSD) value was displayed on the DTS larger than the DTS expected value. It was found that the software incorrectly calculated the PSD value when the x-ray condition of the frontal plane and the lateral plane are the same. It incorrectly used in the calculation a dose value from a previous exposure.
Source facts
- Action
- Toshiba Medical Systems (TAMS) planned action to bring product into Compliance: 1.TAMS will install the software and perform testing to ensure the software update was effective. 2. The customer notification letter which includes a statement that TAMS will, without charge, remedy the defect or bring the product into compliance. . If you have any questions regarding this, please feel free to contact your local Toshiba representative at (800) 521-1968.
- Initiated
- 2016-02-24
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2017-04-25
- Distribution
- US Distribution to the states of : NC, NY, TX, GA, CA, AZ, IL, FL,MA, MO and DE.
- Recall number
- Z-2188-2016
- Recall status
- Terminated
- CDRH record ID
- 147685
- FDA classified
- 2016-07-18
- Recall event number
- 74473