Open original FDA record ↗Recall record
- FDA identifier
- Z-2188-2025
- Company in FDA record
- MICROVENTION INC.
- FDA classification date
- Jul 28, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Drug Elutable Microspheres have a smaller actual average diameter that is not within specification, which may lead to inability to reach the desired treatment location (non-targeted embolization), additional procedure/treatment required, blockage other than target vessel, increased procedure time, incomplete embolization, and/or inability to treat the patient.
Source facts
- Action
- On 5/29/2025, recall notices were mailed and emailed to customers who were asked to do the following: 1) Inform all individuals within your organization of the recall and forward the recall notice to any organizations that may have received the affected products. 2) Complete and return the customer acknowledgement form via email: recalls@microvention.com 3) If your institution has affected inventory call firm's Customer Care at 800-888-3786 or email tmccustomer.admin@terumomedical.com Please direct questions to the firm at MVEMEAQARA@microvention.com
- Initiated
- 2025-05-29
- Root cause
- Device Design
- Distribution
- OUS: Italy, Spain, Portugal, Poland, Belgium, Germany, Denmark, France, Kazakhstan, Slovakia, Norway, Greece, Serbia, Turkey, Czech Republic, United Kingdom, Netherlands, Russian Federation, Kuwait, Republic of North Macedonia, Switzerland, Saudi Arabia, Georgia, Lithuania, Romania, Croatia, Chile, Hong Kong , India, Mexico, Taiwan, Vietnam
- Recall number
- Z-2188-2025
- Recall status
- Open, Classified
- CDRH record ID
- 214022
- FDA classified
- 2025-07-28
- Recall event number
- 96915