Z-2189-2024: Infinity Acute Care System (IACS) Monitoring System, Infinity M500 with Power, Model No. MS20407. Rev 20,

Recall record Z-2189-2024 for Draeger Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2189-2024
Company in FDA record
Draeger Medical Systems, Inc.
FDA classification date
Jun 25, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Device is not in full compliance to Type CF requirements of IEC 60601-1 and IEC 60601-2-34 standards. Therefore, there is a potential risk of electric shock. Health consequences may include electrical shock, burns, and cardiac arrhythmia.

Source facts

Action
On May 13, 2024, the firm began notifying affected customers of the recall through letters titled "Urgent - Medical Device Correction". Customers were informed of the issue and given additional precautions to inspect the patient cables and probes, and to discard damaged or compromised accessories. Customers will be contacted by their local Draeger Service representative to arrange for modification of the affected M500 units to be carried out free of charge as soon as the material for modification is available. If you have any questions regarding the operation of the Product, please contact Drager Service Technical Support between the hours of 8:00 AM - 8:00 PM EST at 1-800-437-2437 (press 2 at the prompt, then 2, then 1). If you have any questions regarding this letter, please contact Michael Kelhart between the hours of 8:00 AM -4:30 PM EST at 267-664-1131 or via email at mike.kelhart@draeger.com.
Initiated
2024-05-13
Root cause
Device Design
Distribution
US Nationwide distribution. International distribution to Afghanistan, Argentina, Armenia, Australia, Austria, Belarus, Belgium, Bolivia, Botswana, Brazil, Brunei Dar-es-S, Bulgaria, Canada, Chile, China, Columbia, Costa Rica, Croatia, Czech Republic, Egypt, Estonia, Finland, France, French Guiana, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Kuwait, Latvia, Liechtenstein, Lithuania, Malaysia, Malta, Martinique, Mexico, Monaco, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, T¿rkiye, Turkmenistan, Ukraine, United Kingdom, Uruguay, United Arab Emirates, Vietnam.
Recall number
Z-2189-2024
Recall status
Open, Classified
CDRH record ID
207919
FDA classified
2024-06-25
Recall event number
94644

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗