- FDA identifier
- Z-2190-2011
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- May 12, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
When using the ARTISTE Linac and RT Therapist v4.1 along with an imported cropped reference image, the image maybe processed incorrectly, thus resulting in a mistreatment.
Source facts
- Action
- Distribution of the Urgent: Medical Device Correction Customer Advisory Notice began on March 24, 2011 via certified mail or by the Siemens Customer Service Group. The letter identified the issue and when it occurs. Users were asked to verify the isocenter position of the reference image in the Portal Review application before using a reference image imported from the treatment planning system. Siemens is currently working on a solution and will notify customers as soon as the update is available. The letter should be included with the Owner Manual of the system in the chapter for "Safety Advisory Letters" until an update is provided.
- Initiated
- 2011-03-24
- Root cause
- Pending
- Terminated
- 2012-02-13
- Distribution
- Worldwide Distribution
- Recall number
- Z-2190-2011
- Recall status
- Terminated
- CDRH record ID
- 99503
- FDA classified
- 2011-05-12
- Recall event number
- 58517