Open original FDA record ↗Recall record
- FDA identifier
- Z-2190-2016
- Company in FDA record
- Synthes (USA) Products LLC
- FDA classification date
- Jul 12, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard.
Source facts
- Action
- DePuy Synthes mailed an URGENT MEDICAL DEVICE FIELD SAFETY NOTIFICATION letter to their customers (Hospital Personnel, including Imaging Department Personnel and DePuy Synthes Colleague) to inform them of the recall and actions to take. The letter identified the affected, problem and actions to be taken. The letter instructed customers not to use the affected Instruments which are labeled or etched MR Safe and MR Conditional during MRI scans. Customers were asked to review the information contained in the Field Safety Notification letter and complete the Verification Section located on page 3 of this notification. For questions call 610-719-5450 or contact your DePuy Synthes Sales Consultant.
- Initiated
- 2016-05-31
- Root cause
- Labeling False and Misleading
- Terminated
- 2017-02-02
- Distribution
- US Nationwide Distribution in the states of: AR,CA,CO,CT,DC,ID,IL,KY,LA,MA,MD,MN,MO,MS,MT,NY,PA and TX.
- Recall number
- Z-2190-2016
- Recall status
- Terminated
- CDRH record ID
- 147472
- FDA classified
- 2016-07-12
- Recall event number
- 74416