Skip to content

Public FDA record

Z-2191-2009: Bard Access Systems, Inc., Fastrac* Guidewire Gastric Access Port Safety System, REF 007035, 24F, Sterile

Recall record Z-2191-2009 for C.R. Bard Access Systems, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2191-2009
Company in FDA record
C.R. Bard Access Systems, Inc
FDA classification date
Sep 15, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Snare can detach in patient. The Polypectomy Snares and the Entake Peg System (which contain these snares) are being recalled because the looped portion of the snare can become detached and drop into the stomach or other internal canal during use, necessitating retrieval of the remaining piece.

Source facts

Action
Bard contacted all 1,883 worldwide customers who received the distributed suspect 467 product code lots, comprising 177, 614 units that were either not expired or were still available in the field. These will be returned by the customer to Bard Access Systems. 1-800-290-1689.
Initiated
2008-12-04
Root cause
Nonconforming Material/Component
Terminated
2011-11-09
Distribution
Worldwide, including USA, Belgium, Canada, Australia, and Japan.
Recall number
Z-2191-2009
Recall status
Terminated
CDRH record ID
84402
FDA classified
2009-09-15
Recall event number
50678

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗