Skip to content

Public FDA record

Z-2192-2014: Gel-E Donut, Squishon 2 gel pillows Product Usage:: The products are intended to support and cradle a

Recall record Z-2192-2014 for Childrens Medical Ventures. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2192-2014
Company in FDA record
Childrens Medical Ventures
FDA classification date
Aug 20, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The gel-filled Gel-E Donut and Squishon product line has received a number of complaints about visible mold. The mold detected was determined to be Cladosporium and Penicillium Fungi, which are commonly found molds.

Source facts

Action
Childrens Medical Ventures sent an Urgent Medical Device Recall letter dated May 28, 2014, to affected customers via UPS. The letter identified the affected product, problem, and actions to be taken. Customers were instructed to review entire inventory and inspect for signs of visible mold, do not open the packaging to do the inspection, the mold should be seen through the transparent outer packing, complete and return the reply form provided on the last page of the letter, dispose of any product containing mold per your facilitys environmental guidelines, do not return any product to your Distributor or to Philips/Childrens Medical Ventures.and credit for disposed product will be given. If you need any further information or support concerning this problem, please contact your Distributor at the following phone number:
Initiated
2014-05-28
Root cause
Material/Component Contamination
Terminated
2016-11-29
Distribution
Worldwide Distribution - USA (nationwide) including the country of Canada.
Recall number
Z-2192-2014
Recall status
Terminated
CDRH record ID
127715
FDA classified
2014-08-20
Recall event number
68443

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗