Open original FDA record ↗Recall record
- FDA identifier
- Z-2192-2014
- Company in FDA record
- Childrens Medical Ventures
- FDA classification date
- Aug 20, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
The gel-filled Gel-E Donut and Squishon product line has received a number of complaints about visible mold. The mold detected was determined to be Cladosporium and Penicillium Fungi, which are commonly found molds.
Source facts
- Action
- Childrens Medical Ventures sent an Urgent Medical Device Recall letter dated May 28, 2014, to affected customers via UPS. The letter identified the affected product, problem, and actions to be taken. Customers were instructed to review entire inventory and inspect for signs of visible mold, do not open the packaging to do the inspection, the mold should be seen through the transparent outer packing, complete and return the reply form provided on the last page of the letter, dispose of any product containing mold per your facilitys environmental guidelines, do not return any product to your Distributor or to Philips/Childrens Medical Ventures.and credit for disposed product will be given. If you need any further information or support concerning this problem, please contact your Distributor at the following phone number:
- Initiated
- 2014-05-28
- Root cause
- Material/Component Contamination
- Terminated
- 2016-11-29
- Distribution
- Worldwide Distribution - USA (nationwide) including the country of Canada.
- Recall number
- Z-2192-2014
- Recall status
- Terminated
- CDRH record ID
- 127715
- FDA classified
- 2014-08-20
- Recall event number
- 68443