Public FDA record

Z-2192-2020: Power supply provided with Afinion 2, Catalog numbers 1116553, 1116554, 1116556, 1116557, 1116597, 111659

Recall record Z-2192-2020 for Abbott Diagnostics Technologies AS. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2192-2020
Company in FDA record
Abbott Diagnostics Technologies AS
FDA record creation date
May 29, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The analyzer system may have an electrostatic discharge and sparking of the power supply provided with the analyzers resulting in the power supply becoming non-functional.

Source facts

Action
The recalling firm/manufacturer issued medical device correction letters dated 3/23/2020 on 3/23/2020 to their direct distributors requesting end user customer information. The recalling firm/manufacturer issued medical device correction letters to the end users dated 3/31/2020 via a third-party notification firm. The notification informed the end user of the issue and requested to be notified of the number of power supplies needed for replacement. The affected power supplies were to be discarded by the end user.
Initiated
2020-03-23
Root cause
Device Design
Distribution
Distribution was made to NJ. There was no direct government/military distribution. Foreign distribution was made to Canada, Argentina, Australia, Bangladesh, Brazil, Cambodia, Chile, China, Colombia, Denmark, Finland, France, Georgia, Germany, Guatemala, Hong Kong, Iceland, India, Italy, Ivory Coast, Japan, Korea, Madagascar, Malaysia, Mexico, Mongolia, Myanmar, The Netherlands, Norway, Oman, Pakistan, Peru, Philippines, Poland, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Timur, Trinidad & Tobago, United Kingdom, and Zimbabwe.
Recall number
Z-2192-2020
Recall status
Open, Classified
CDRH record ID
180494
FDA record created
2020-05-29
Recall event number
85313

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗