Skip to content

Public FDA record

Z-2193-2008: AMPLATZER Sizing Balloon II, REF 9-SB-24. Sterile EO. AGA Medical Corporation, 682 Mendelssohn Avenue,

Recall record Z-2193-2008 for AGA Medical Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2193-2008
Company in FDA record
AGA Medical Corporation
FDA classification date
Sep 3, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

AGA Medical Corporation has identified an issue with outer pouch seal integrity within one production lot of 10 French 45/80 delivery systems and one lot of 24 mm Sizing Balloon. We have not observed incomplete seals with the inner pouch. An intact inner pouch seal will maintain device sterility. AGA is asking that customers inspect the specific lot of devices to confirm the integrity of the ou

Source facts

Action
The firm Fed Ex'd and emailed a Field Inspection/Correction letter to customer 04/28/2008. The letter identified affected product, requested that customers check the outer package seal, and if it is incomplete the device should be returned for replacement. A form was also asked to be returned to AGA Medical. Contact AGA Medical at 1-763-513-9227 for assistance.
Initiated
2008-04-25
Root cause
Packaging process control
Terminated
2010-10-13
Distribution
Within the US to include: AL, AZ, CA, CT, FL, IA, ID, MA, MD,MI, MO, NC, NH, NY, OH, PA, SD, TN, TX, WI. OUS to include: Brazil, Greece, Netherlands, Portugal, and Slovakia
Recall number
Z-2193-2008
Recall status
Terminated
CDRH record ID
71806
FDA classified
2008-09-03
Recall event number
48650

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗