Open original FDA record ↗Recall record
- FDA identifier
- Z-2193-2013
- Company in FDA record
- Carefusion 2200 Inc
- FDA classification date
- Sep 11, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Catheters included in Thoracentesis Tray and Safe-T Thoracentesis Tray - Drug Free may occlude, preventing drainage of excessive fluid from the pleural cavity leading to a potential delay in the medical procedure and potentially serious patient complications.
Source facts
- Action
- Recall notification letters dated 8/27/13 were distributed via courier to direct accounts. Consignees were/are requested to confirm receipt of the notification, examine their inventory for the recalled products and to destroy the products. After the products are destroyed, the consignees are requested to submit a completed response form to CareFusion for replacement with functional devices. Consignees who cannot destroy the affected products are requested to contact CareFusion for additional information.
- Initiated
- 2013-08-28
- Root cause
- Process change control
- Terminated
- 2015-06-16
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MD, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, and WV and Puerto Rico, and the country of Iceland.
- Recall number
- Z-2193-2013
- Recall status
- Terminated
- CDRH record ID
- 121524
- FDA classified
- 2013-09-11
- Recall event number
- 66123