Skip to content

Public FDA record

Z-2193-2014: Prismaflex TPE 2000 set, Product Number 1071441. Each set is packaged in a plastic pouch, 4 pouches per c

Recall record Z-2193-2014 for Gambro Renal Products, Incorporated. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2193-2014
Company in FDA record
Gambro Renal Products, Incorporated
FDA classification date
Aug 13, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Gambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood.

Source facts

Action
Gambro sent an Urgent Field Safety Notice dated July 18, 2014 to all end-users. A Customer Reply Form was attached to the notice for customers to complete and return.
Initiated
2014-07-18
Root cause
Device Design
Terminated
2015-02-05
Distribution
Worldwide Distribution.
Recall number
Z-2193-2014
Recall status
Terminated
CDRH record ID
128667
FDA classified
2014-08-13
Recall event number
68788

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗