Z-2194-2010: The Philips EasyDiagnost Eleva with DRF. Philips Medical Systems multifunctional x-ray system.

Recall record Z-2194-2010 for Philips Medical Systems North America Co. Phillips. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2194-2010
Company in FDA record
Philips Medical Systems North America Co. Phillips
FDA classification date
Aug 10, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Potential for bright artifacts to appear on the patient image from a previous exposure.

Source facts

Action
On July 9, 2008, Philip Medical System began sending out the URGENT-Device Correction letter to their consignees. This letter describes the recalled product as EasyDiagnost Eleva with DRF system with SW release 1.0. The firm informs the consignees that under special circumstances there is a small possibility that bright artifacts appear on a patient image from a previous exposure. The consignees are advised to take care about the X-ray field limitation. If the detector is not affected by direct radiation, no problem will occur. If direct radiation reaches the detector, the collimation should not be too tight, such that the area of direct radiation is larger than 1.2 inch x 1.2 inch (or 30 mm x 30 mm). The minimum area is needed to detect and prevent memory artifacts. The firm will upgrade any affected products with the Field Change Order (FCO)70600031 and it will be implemented before the end of December 2008. The upgrade software is 1.0.3. The correction would be free of charge and the consignees should contact Philips Call Center at 800-772-9377, #5, #2.
Initiated
2008-07-09
Root cause
Other
Terminated
2010-10-27
Distribution
Product was distributed in AZ, CA, FL and WA.
Recall number
Z-2194-2010
Recall status
Terminated
CDRH record ID
72491
FDA classified
2010-08-10
Recall event number
48998

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗