Skip to content

Public FDA record

Z-2194-2020: Gambro Cartridge Blood Transport System for Hemodialysis Blood Set - Product Usage: is intended for singl

Recall record Z-2194-2020 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2194-2020
Company in FDA record
Baxter Healthcare Corporation
FDA record creation date
Jun 1, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential disconnection of tubing set.

Source facts

Action
Baxter Healthcare notified customers on about 04/17/2020 via "Urgent Device Correction" letter. Customers were instructed to do the following: 1. Prior to use, thoroughly inspect each connection of the set to check for any detachments in the tubing. Operators may continue to use affected sets if no detachments are observed. Additionally, per the Instructions For Use (IFU), users should observe carefully for leaks during priming and use, and examine the tubing carefully to be certain that all connections are secure, all lines are unobstructed and that there are no kinks or leaks in the tubing. 2. If you find sets with tubing disconnections, please contact Baxter Healthcare Center for Service to arrange for return and credit. Baxter Healthcare Center for Service can be reached at 888-229-0001 between the hours of 7:00 am and 6:00 pm Central Time, Monday through Friday. Please have your Baxter 8-digit ship-to account number, product code, lot number, and quantity of product to be returned ready when calling. 3. If you purchased affected product directly from Baxter, a customer reply form is included in your mailing. Please complete the enclosed Baxter customer reply form and return it to Baxter by faxing it to 224-270-5457 or scanning and e-mailing it to fca@baxter.com, even if you do not have any inventory. Returning the customer reply form promptly will confirm your receipt of this notification and prevent you from receiving repeat notices. 4. If you purchased this product from a distributor, please note that the Baxter customer reply form is not applicable. If a reply form is provided by your distributor or wholesaler, please return it to the supplier according to their instructions. 5. If you distribute this product to other facilities or departments within your institution, please forward a copy of this communication to them. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that
Initiated
2020-04-17
Root cause
Process control
Terminated
2023-02-07
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY. Countries of Australia, Austria, Bahrain, Belgium, Canada, Denmark, Egypt, Finland, France, Germany, Greece, Israel, Italy, Kuwait, Lithuania, Latvia, Montenegro, Morocco, Netherlands, Norway, Poland, Russia, Serbia, Slovenia, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom, Korea, Singapore, Hong Kong
Recall number
Z-2194-2020
Recall status
Terminated
CDRH record ID
181241
FDA record created
2020-06-01
Recall event number
85514

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗