The firm sent an Update Instruction, "TH003/08/S" for Motion Enable Switch/holder assembly. The switch holder allows the motion enable switch to be activated if inserted partially; thereby enabling inadvertent motion and serious injury may result.
Source facts
Action
On June 25, 2008 all consignees were notified via written letter and on-site visit by the firm's sales representative, informing them that the switch holder allows the motion enable switch to be activated if inserted partially; thereby enabling inadvertent motion and serious injury may result and providing instructions on the recall. A new software version will be provided. For assistance contact Siemens at 1-925-246-8407.
Initiated
2008-03-10
Root cause
Software design
Terminated
2010-12-22
Distribution
Worldwide and Nationwide
Recall number
Z-2196-2008
Recall status
Terminated
CDRH record ID
71815
FDA classified
2008-06-28
Recall event number
48655