Skip to content

Public FDA record

Z-2196-2010: ASCENT HEALTHCARE SOLUTIONS. Reprocessed Femoral Compression Device, Model 11165.

Recall record Z-2196-2010 for Ascent Healthcare Solutions, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2196-2010
Company in FDA record
Ascent Healthcare Solutions, Inc.
FDA classification date
Aug 10, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The FemoStop Devices may fail to inflate or hold pressure. A separation between the dome and arch base prevents the device from maintaining pressure when inflated.

Source facts

Action
Ascent sent an urgent medical device voluntary recall notification Letter and Response Form to each consignee on June 25, 2010. Consignees were requested to immediately examine their inventory of Ascent Reprocessed FemoStop devices with lot numbers located on an attached list; remove and quarantine all affected devices; complete the attached Recall Effectiveness Check Form; ship all affected product back to Ascent using the address indicated in the letter, and fax a completed Recall Effectiveness Check Form to 480-763-5350.
Initiated
2010-06-23
Root cause
Reprocessing Controls
Terminated
2010-08-11
Distribution
Devices were distributed to 8 consignees (end users/hospitals) located in VA, CA, FL, MI, WA.
Recall number
Z-2196-2010
Recall status
Terminated
CDRH record ID
92761
FDA classified
2010-08-10
Recall event number
56157

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗