Open original FDA record ↗Recall record
- FDA identifier
- Z-2196-2010
- Company in FDA record
- Ascent Healthcare Solutions, Inc.
- FDA classification date
- Aug 10, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The FemoStop Devices may fail to inflate or hold pressure. A separation between the dome and arch base prevents the device from maintaining pressure when inflated.
Source facts
- Action
- Ascent sent an urgent medical device voluntary recall notification Letter and Response Form to each consignee on June 25, 2010. Consignees were requested to immediately examine their inventory of Ascent Reprocessed FemoStop devices with lot numbers located on an attached list; remove and quarantine all affected devices; complete the attached Recall Effectiveness Check Form; ship all affected product back to Ascent using the address indicated in the letter, and fax a completed Recall Effectiveness Check Form to 480-763-5350.
- Initiated
- 2010-06-23
- Root cause
- Reprocessing Controls
- Terminated
- 2010-08-11
- Distribution
- Devices were distributed to 8 consignees (end users/hospitals) located in VA, CA, FL, MI, WA.
- Recall number
- Z-2196-2010
- Recall status
- Terminated
- CDRH record ID
- 92761
- FDA classified
- 2010-08-10
- Recall event number
- 56157