Skip to content

Public FDA record

Z-2197-2020: OLYMPUS ShockPulse-SE Lithotripsy System, Model SPL-T Transducer

Recall record Z-2197-2020 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2197-2020
Company in FDA record
Olympus Corporation of the Americas
FDA record creation date
Jun 1, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A component of the Olympus ShockPulse Transducer was manufactured with non-conforming material that could allow water ingress during autoclaving sterilization.

Source facts

Action
Olympus sent a Medical Device Field Removal Action letter, dated April 23, 2020, to all ShockPulse Transducer (Model SPL-T) customers with the affected serial numbers. The letter states that the transducer was manufactured with non-conforming material and could become warped during sterilization. They explained that the serial number of the transducers is found etched on the back of the device. The firm asked that the customers determine whether they have any of the affected product in stock. If customers have any of the affected product , they were instructed to quarantine the product. and contact the Olympus customer service representative at 1-888-524-7266 option 1, in order to receive a Return Material Authorization for any affected product then, get issued a credit or replacement. Customers were also instructed to access the Olympus recall portal (https://olympusamerica.com/recall) to indicate that they received the notification. If they distributed the product further, the firm asked them to notify their customers to inform them of the field action.
Initiated
2020-04-23
Root cause
Component design/selection
Terminated
2024-10-09
Distribution
Distribution US nationwide to states of: Texas, Ohio, New Mexico, California, Oklahoma, Michigan, Connecticut, North Carolina, Nebraska, and Pennsylvania; and Canada.
Recall number
Z-2197-2020
Recall status
Terminated
CDRH record ID
181234
FDA record created
2020-06-01
Recall event number
85558

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗