Open original FDA record ↗Recall record
- FDA identifier
- Z-2198-2008
- Company in FDA record
- Greiner Bio-One North America, Inc.
- FDA classification date
- Sep 30, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
At removal of the tube from the holder the needle may dislodge and become stuck in the tube stopper with the blunt end of the needle facing out.
Source facts
- Action
- Consignees were notified by letter on 08/06/2008. Letters to Distributors instructed them to stop distribution of the affected product immediately and to hold any product of the lots they may have pending returned to Greiner bio-one. The were instructed to provide Greiner bio-one with a list of their customers who may have received the affected lot so that they can contact them. A Product Disposition form was included to be completed and returned to Greiner bio-one. Contact Greiner Bio-One North America, Inc. at 1-888-286-3883 for assistance.
- Initiated
- 2008-08-05
- Root cause
- Process design
- Terminated
- 2012-07-25
- Distribution
- AL, FL, IN, KS, KY, MO, MN, NC, OH, PA, TN, TX, UT, VA, WI and WV.
- Recall number
- Z-2198-2008
- Recall status
- Terminated
- CDRH record ID
- 73213
- FDA classified
- 2008-09-30
- Recall event number
- 49367