Skip to content

Public FDA record

Z-2198-2010: CERCLAGE WIRE, 0.8 MM DIA. X 200 MM LENGTH, SST, REF 71163008, QTY: (6), STERILE R, CE 008, 2020/07, Smi

Recall record Z-2198-2010 for Smith & Nephew Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2198-2010
Company in FDA record
Smith & Nephew Inc
FDA classification date
Aug 10, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Plastic trays which encase these products have burrs on the hinged packaging which can puncture the single barrier sterile pouch, compromising sterility.

Source facts

Action
All affected Smith & Nephew Sales Representatives were notified of problem and the recall was sent via overnight mail on 01/23/2009. All affected Smith & Nephew International Distributors were notified via email and telephone on 01/23/2009. They were instructed to immediately quarantine the product for return.
Initiated
2010-06-28
Root cause
Process control
Terminated
2012-01-23
Distribution
KY, TX, DC, NC, CA, AZ, GA and OH; 3 International: Australia, Malaysia, and Switzerland
Recall number
Z-2198-2010
Recall status
Terminated
CDRH record ID
92888
FDA classified
2010-08-10
Recall event number
56187

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗