Z-2198-2020: FlexLab Augomation System using the Aliquoter Module (Inpeco P/N FLX-212) The systems consolidate multipl

Recall record Z-2198-2020 for Inpeco S.A.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2198-2020
Company in FDA record
Inpeco S.A.
FDA record creation date
Jun 1, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for discrepant patient results due to a potential for the primary sample tube to be diluted. In certain firmware versions, in case a Clot Detection error (error code E0E0 or 13E0) is generated during the sample aspiration the current error recovery procedure dispenses 2/3 of sample volume back into the Primary Tube. Evidence from the field showed that in case of Clot Detection error, this management may lead to the dilution of the Primary Tube with the distilled water of the hydraulic circuit of the Aliquoter Module.

Source facts

Action
Urgent: Medical Device Correction dated 03/30/2020 were sent to customers. The impacted modules are the Aliquoter Modules (Inpeco Part Number FLX-212) with a firmware version prior to the following: AQMb_3-3-0.H86; AQMa_3-1-1-8.H86 and AQMb_3-1-1-8.H86; xAQMb_1-1-0.elf; Action to be taken by the user To avoid the risk of contamination take the following precaution: 1) Discard the Primary Tube flagged with Clot Detection error or manage it according to your laboratory guidelines considering that it may be diluted. 2) Call Service Assistant in case the frequency of the Clot Detection Error increases (more than 5 consecutive Clot Detection Errors). Your service provider will contact you to schedule the firmware upgrade. Until the service visit please maintain awareness on this notice. Please complete and return the "Customer Letter Receipt Confirmation and Implementation CHeck form" attached to this letter with 15 days directly at the email address specified in the email communication. Contact reference person: FOr any clarificatino you may need, do not hesitate to ocntact: Eva Balzarotti - REgulatory Affairs Manager E-mail: regulatory.affairs@inpeco.com Phone: (+41) 91 9118 224
Initiated
2020-04-07
Root cause
Software design
Distribution
US Nationwide distribution including in the states of IL, NY.
Recall number
Z-2198-2020
Recall status
Open, Classified
CDRH record ID
181021
FDA record created
2020-06-01
Recall event number
85469

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗