Sampling set nasal cannula used with End-Tidal Carbon Monoxide (ETCO) monitor may have residual carbon monoxide and volatile organic compounds present due to insufficient aeration of the cannula's carbon filter prior to sealing, which may lead to higher than expected ETOCc scores during clinical evaluations.
Source facts
Action
Starting on 4/21/2025 Product Advisory and Voluntary Field Action emails were sent to customers who were told affected devices would be replaced. Customers with questions could contact the firm at 904-588-2986
Initiated
2025-04-21
Root cause
Device Design
Distribution
US Nationwide distribution in the states of PA, UT, TX, FL, MA, IA.
Recall number
Z-2198-2025
Recall status
Open, Classified
CDRH record ID
214760
FDA classified
2025-07-29
Recall event number
97184