Open original FDA record ↗Recall record
- FDA identifier
- Z-2199-2019
- Company in FDA record
- Philips Ultrasound Inc
- FDA classification date
- Aug 6, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a potential that the image of one patient could get unexpectedly moved into another patient's folder when the user is using the Edit functionality on the Patient Data Entry (PDE) screen of the device.
Source facts
- Action
- The recalling firm issued letters dated 4/26/2019 on 4/29/2019 via certified mail with delivery confirmation. The letter identified the affected product, problem and actions to be taken.
- Initiated
- 2019-04-29
- Root cause
- Software design
- Terminated
- 2022-08-04
- Distribution
- Worldwide Distribution - US Nationwide, including Puerto Rico. There was also military/government distribution. Foreign distribution was made to Canada, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Israel, Italy, Japan, Latvia, Luxembourg, Malaysia, Mayotte, Mexico, Monaco, Morocco, Netherlands, New Zealand, Nicaragua, Norway, Oman, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Russian Federation, Saudi Arabia, Serbia, Singapore, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, United Kingdom Virgin Islands, Uruguay, and Viet Nam.
- Recall number
- Z-2199-2019
- Recall status
- Terminated
- CDRH record ID
- 173308
- FDA classified
- 2019-08-06
- Recall event number
- 83003