- FDA identifier
- Z-2200-2013
- Company in FDA record
- Medtronic Sofamor Danek USA Inc
- FDA classification date
- Sep 14, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
One lot of the T8 Removal Head had incorrect hardness specification, which could result in the T8 Removal head failing during removal of the CD Horizon Spire and Spire Z Spinal System implant. There is a potential for tissue damage to occur and/or minor surgical delay to remove the remaining fragments of the tip.
Source facts
- Action
- Medtronic sent an Urgent Voluntary Market Removal Recall letter dated August 9, 2013 with Recall Questionnaires and Confirmation forms included to Risk Managers and Sales Representatives. The letter identified the affected product, problem and action to be taken. Customers were instructed to remove the affected product, discontinue use and contact your Medtronic Sales Representative to coordinate the return and replenishment of affected product. For questions call your Medtronic Sales Representative or Global Quality Department at 800-876-3133.
- Initiated
- 2013-08-12
- Root cause
- Nonconforming Material/Component
- Terminated
- 2013-12-27
- Distribution
- USA Nationwide Distribution in the state of GA, AK, WA, and CA.
- Recall number
- Z-2200-2013
- Recall status
- Terminated
- CDRH record ID
- 121220
- FDA classified
- 2013-09-14
- Recall event number
- 66010