Open original FDA record ↗Recall record
- FDA identifier
- Z-2200-2026
- Company in FDA record
- MICROVENTION INC.
- FDA classification date
- May 14, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Stent-assisted coiling device may due to commingling during manufacturing have dimensional nonconformance and/or incorrect device configuration. If device is oversized may lead to increased vessel wall stress, neointimal hyperplasia, vessel injury/rupture; if undersized may lead to incomplete expansion/malposition, acute stent thrombosis, distal embolization, ischemic neurological events
Source facts
- Action
- On 4/6/2026, recall notices with TERUMO NEURO listed on notice top, were emailed to customers who were asked to do the following: 1) Stop using and quarantine affected devices. 2) Complete and return the acknowledgement form via email to -Distributors should provide the recall notice to medical facilities or users to whom you have distributed affected products. - Distributors should reconcile the completed forms from all customers and submit to recalling firm via email mvchinaqa@microvention.com If you have questions about this recall contact firm at max.huang@microvention.com
- Initiated
- 2026-04-06
- Root cause
- Mixed-up of materials/components
- Distribution
- International distribution to the country of China.
- Recall number
- Z-2200-2026
- Recall status
- Open, Classified
- CDRH record ID
- 219701
- FDA classified
- 2026-05-14
- Recall event number
- 98755