Open original FDA record ↗Recall record
- FDA identifier
- Z-2201-2012
- Company in FDA record
- Nico Corp.
- FDA classification date
- Aug 13, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
BrainPath sheaths were mislabeled. Sheaths which measure 50mm were labeled as 60mm. The length callout on the label does not match the actual length of the sheath within the packaging resulting in an improper fit as further explained below. This mislabeling was discovered on 29-June-2012 during a clinical case.
Source facts
- Action
- NICO Neuro and Spine sent an Urgent Recall letter to all affected customers on July 18, 2012. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to complete the recall response form. Affected sheaths distributed within the US will be obtained and discarded by NICO. Affected sheaths shipped to Canada will be obtained and discarded by NICOs distributor, Galen Medical. For questions customers should call 888-632-7071.
- Initiated
- 2012-07-17
- Root cause
- Labeling mix-ups
- Terminated
- 2012-11-07
- Distribution
- Worldwide Distribution - USA including MI and Canada
- Recall number
- Z-2201-2012
- Recall status
- Terminated
- CDRH record ID
- 111099
- FDA classified
- 2012-08-13
- Recall event number
- 62588