Open original FDA record ↗Recall record
- FDA identifier
- Z-2202-2015
- Company in FDA record
- Abbott Laboratories
- FDA classification date
- Jul 22, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The ARCHITECT c8000 instrument contains tubing that does not meet specifications and has the potential to leak.
Source facts
- Action
- Abbott notified thier affected consignees via phone or visit on 5/29/15. A follow up Product Recall notification letter dated June 4, 2015 was sent to affected customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to immediately discontinue use of the product, Abbott representative will be make all necessary arrangements for replacing your impacted instruments(s), follow your laboratory protocol regarding the need to review previously reported patient results and retain this notification for laboratory records. For questions contact Customer Service at 1-877-4ABBOTT.
- Initiated
- 2015-06-11
- Root cause
- Mixed-up of materials/components
- Terminated
- 2016-02-09
- Distribution
- US Nationwide Distribution in the states of CA, AL, NY and WI.
- Recall number
- Z-2202-2015
- Recall status
- Terminated
- CDRH record ID
- 137918
- FDA classified
- 2015-07-22
- Recall event number
- 71473