Under specific conditions (Fluoroscopic exam, ABS=OFF, manual change of parameters) wrong calculation of the dose/minute for fluoroscopy exams can be possible
Source facts
Action
Agfa issued Urgent Field Safety Notice via email on 4/21/20 to consignees, letter states reason for recall, health risk and action to take: Acknowledgment, via a feedback form or email, that the information was received and understood was requested from the consignees.
Initiated
2020-04-06
Root cause
Software design
Terminated
2023-04-03
Distribution
US Nationwide distribution including in the sates of FL, IL, IN, LA, MA, MO, OH, SC, VA, WI.
Recall number
Z-2202-2020
Recall status
Terminated
CDRH record ID
181310
FDA record created
2020-06-03
Recall event number
85599