Reason stated in the source
Software: Audible alarm may be silenced if changed from the factory setting
Source facts
- Action
- CASMED notified customers by a Product Correction Notification letter on 6/24/08 to advise of problem of the audible alarm and a software upgrade to correct the problem. The firm requests that the alarm volume not be adjusted from its factory set value to ensure that alarms sound correctly. Also, the letter states that if the volume needs to be adjusted, that customers needs to be sure that they only select two clicks above, or two clicks below the mid-point, and that there is a corresponding beep when you make your selection. Customers are asked to review the attached list of serial numbers that will need to be updated with new software and to then email the a contact at the firm once updates have been completed. All updates must be completed by August 31, 2008.
- Initiated
- 2008-06-24
- Root cause
- Software change control
- Terminated
- 2010-08-05
- Distribution
- Worldwide Distribution --- including USA and countries of Canada, Jordan, UK, France, and Taipei.
- Recall number
- Z-2203-2008
- Recall status
- Terminated
- CDRH record ID
- 71838
- FDA classified
- 2008-09-16
- Recall event number
- 48666