Z-2203-2020: Pressure Monitoring Set, C-PMS, G02838

Recall record Z-2203-2020 for Cook Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2203-2020
Company in FDA record
Cook Inc.
FDA classification date
Jun 2, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Pressure Monitoring Set and Tray may contain excess coating material on the outside and/or interior of the needle component which could cause occlusion.

Source facts

Action
On April 30, 2020, Cook Medical sent an Urgent Medical Device Recall letter to their customers to inform them of the voluntary recall of specific lots of the Pressure Monitoring Set and Pressure Monitoring Tray. The firm instructed customers to examine their inventory immediately to determine if have affected product(s) and quarantine any affected product that remains unused then, cease all distribution and use of the product. Cook Medical requested that customers complete the Acknowledgement and Receipt Form via fax or email within 5 business days of receiving the recall letter, whether or not they had any affected product(s) on hand. Customers were instructed to return the affected product(s) with a copy of the Acknowledgement and Receipt Form to receive a product credit. ***Updated 7/7/20*** The firm sent an updated letter dated July 7, 2020, to inform customers of 14 additional lots of the Pressure Monitoring Tray (GPN G02854) that were inadvertently excluded from the previous letter.
Initiated
2020-04-30
Root cause
Process control
Terminated
2021-03-29
Distribution
Worldwide distribution including distribution to the following states: Alaska, Alabama, Arkansas, Arizona, California, Colorado, Connecticut, Washington DC, Delaware, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, North Carolina, North Dakota, Nebraska, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, Wisconsin, and West Virginia
Recall number
Z-2203-2020
Recall status
Terminated
CDRH record ID
181329
FDA classified
2020-06-02
Recall event number
85605

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗