Open original FDA record ↗Recall record
- FDA identifier
- Z-2204-2008
- Company in FDA record
- Intuitive Surgical, Inc.
- FDA classification date
- Sep 16, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Defective software chip may cause the system to fail and lock up.
Source facts
- Action
- Recall initiated June 1, 2008. Urgent Device Recall Letters were sent via e-mail, with follow up phone calls and service visits to correct the product. The firm recommended that continued use of the da Vinci S Surgical System during the time period prior to our service update to replace the software chip. The letter emphasizes the importance of always having backup equipment and instrumentation available when performing surgery with the da Vinci S Surgical System and being prepared to convert to alternative surgical techniques, if necessary. Customers are asked to fill out and return the acknowledgement form.
- Initiated
- 2008-06-01
- Root cause
- Software design (manufacturing process)
- Terminated
- 2009-07-29
- Distribution
- Worldwide Distribution --- including USA and countries of Belgium, Bulgaria, Canada, China, Denmark, France, Germany, Italy, Korea, Romania, Russia, Saudi Arabia, Singapore, Spain, Sweden, and United Kingdom.
- Recall number
- Z-2204-2008
- Recall status
- Terminated
- CDRH record ID
- 71855
- FDA classified
- 2008-09-16
- Recall event number
- 48680