On June 12, 2019 Ad-Tech was made aware that on three different occasions between March 1, 2019 and May 31, 2019 Ad-Tech personnel have distributed non-sterile anchor bolt parts to one independent representative and one international distributor, who may have subsequently sterilized the non-sterile anchor bolts for clinical use. Ad-Tech does not have regulatory clearance to distribute non-sterile anchor bolts and does not have a validated hospital sterilization method for these devices. Further, Ad-Tech labeling does not include instructions for sterilizing anchor bolts.
Source facts
Action
Ad-Tech Medical notified customers on about 06/18/2019 via "URGENT: MEDICAL DEVICE RECALL" letter. Customers were instructed to immediately examine inventory and quarantine any parts subject to the recall, notify all customers if the affected parts have been further distributed, and provide a copy of the recall notification letter. Additionally, customers were instructed to complete and return the customer response form, contact Ad-Tech Customer Support Specialists to obtain a Return Material Authorization (RMA) number, and return all affected parts to Ad-Tech Medical Instrument Corporation at 400 West Oakview Parkway, Oak Creek, WI 53154.
The customer response form should be sent to the Ad-Tech Regulatory Team at FAX: 262-634-5668, Telephone: 262-634-1555, Email: Regulatory@adtechmedical.com. Questions and concerns can also be directed to the Ad-Tech Regulatory Team at the above contact information Monday through Friday, 7:00 AM to 5:00 PM, Central Time.
Initiated
2019-06-18
Root cause
Employee error
Terminated
2021-02-16
Distribution
Worldwide distribution - US Nationwide distribution in the state of Georgia and country of Australia.
Recall number
Z-2204-2019
Recall status
Terminated
CDRH record ID
174373
FDA record created
2019-08-06
Recall event number
83246