Z-2204-2021: Alaris Infusion Pump Module, Model 8100, serviced/repaired by Biomedical Equipment Service Co. (BMES) wit

Recall record Z-2204-2021 for BioMedical Equipment Service Co (BMES). View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2204-2021
Company in FDA record
BioMedical Equipment Service Co (BMES)
FDA classification date
Aug 6, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.

Source facts

Action
Recall notification emails were sent to customers beginning 2/25/21. The emails stated the following: We have been notified by the FDA and our distributor about a voluntary part recall, to address potential performance concerns for a bezel part used in the service and repair of Alaris Pump Module 8100. The bezel part subject to this corrective action was used in the service and repair of the Alaris Pump Module 8100 between July 10, 2020 and February 18, 2021. The part is from a distributor and is not a BMES part. We have switched distributors since learning of the issue. While it's listed as a voluntary recall we do not consider fixing the issue voluntary. BMES will cover the replacement of all impacted bezel repair parts under this field safety corrective action. A detailed letter is attached with the specifics outlined below. Formal Letter Customer Response Form Images of Impacted bezel repair part List of Impacted repairs. Our records indicate that BMES performed service and repair, including the installation of the impacted part on (X) out of a quantity of (Y) Alaris 8100 repairs. Next Steps REad through the attached letter and complete the attached form titled attachment A within the attached letter and email it to our BMES Quality Assurance, Jesse Curry at jcurry@bmesco.com. After placing the potentially affected pump in quarantine, please contact Quality Assurance at 1-800-626-4515 or email Jesse Curry at jcurry@bmesco.com to schedule replacement of the impacted part. The letter attached to the emails, titled "URGENT: Voluntary Field Safety Corrective Action Bezel Repair Part for Alaris Pump Module Model 8100" provided additional details regarding the product issue and the associated health risks.
Initiated
2021-02-25
Root cause
Component design/selection
Terminated
2024-04-22
Distribution
The affected products were distributed to the following US states: CA, IL, IN, KY, LA, NJ, NV, OH, and PA.
Recall number
Z-2204-2021
Recall status
Terminated
CDRH record ID
188034
FDA classified
2021-08-06
Recall event number
88170

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗