Z-2205-2019: Philips Azurion systems with software version R1.2 -Interventional Fluoroscopic X-ray system as follows

Recall record Z-2205-2019 for Philips North America, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2205-2019
Company in FDA record
Philips North America, LLC
FDA record creation date
Aug 6, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The cold restart of Azurion R1.2 systems may take up to 7 minutes if the system is connected to the mains power supply for more than 50 days, may result in a delay of treatment.

Source facts

Action
Philips issued notification on July 16, 2019 advising users of the problem, health risk and action to take: If the start-up screen shows software version R1 .2, your system has the affected release and FC072200430 should be implemented on your system. If the start-up screen shows software version R1.2.1, your system has the latest system software release and is not affected by this issue. Philips will also be contacting customers with affected systems directly. Until Philips corrects your Philips Azurion system. At least once every 50 days, briefly disconnect the Azurion system from the mains power supply. Be aware that when an Uninterruptible Power Supply (UPS) is installed, the room emergency power off switch should be used to interrupt the mains power supply to the system. Note that this might also shut down other devices or equipment in the room. . Include this Field Safety Notice with the documentation of the system until Philips implements this correction in your system. Questions contact : your local Philips representative
Initiated
2019-07-16
Root cause
Software design
Terminated
2020-06-04
Distribution
US Nationwide Distribution Foreign: Canada Argentina AUSTRALIA Austria Azerbaijan Bahrain Bangladesh Belgium Bosnia and Herzegovina Brazil Bulgaria Chile Colombia CROATIA Denmark ECUADOR Egypt Finland France Germany Greece Hong Kong India Indonesia Ireland Israel Italie Italy Japan Jordan Kazakhstan Korea, Republic of Latvia Lebanon Libya Malaysia Maldives Mexico Mongolia Myanmar Netherlands New Zealand Norway Oman Pakistan Philippines Polamnd Poland Portugal Republic of Korea Romania Russian Federation Saudi Arabia Serbia Singapore SLOVENIA South Africa Spain Sri Lanka Sweden Switzerland Thailand Tunisia Turkey UNITED ARAB EMI United Arab Emirates United Kingdom Uzbekistan Viet Nam
Recall number
Z-2205-2019
Recall status
Terminated
CDRH record ID
175362
FDA record created
2019-08-06
Recall event number
83477

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗